Technical data
|
Energy range | | 2.3 – 12 keV | | Resolution | | ΔE from 100 meV to 1 eV at 8 keV | Source | | In-vacuum U20 undulator | | Polarization | | linear | | Spot size | | 20 (V) x 80 (H) µm2 in standard mode, 5 (V) x 10 (H) µm2 in KB configuration | | Optics | | 2-bounce fixed-exit monochromator (Si 111); 4-bounce high-resolution monochromator (Si 311 et 220); collimating and focusing mirrors; 2 mirrors in KB configuration | | Flux | | 1.5x1012 ph/s (standard foc.), 5x1011 (micro-foc.) at 8 keV | Detectors | | Si diodes | | Instrument | | 2-m radius IXS spectrometer (horizontal and vertical scattering), 0.5 m radius high-flux spectrometer; high energy / high resolution electron analyzer (30 meV à 12 keV) |
|
Scientific Opportunities
|
| Hard condensed matter | | Strongly correlated electrons, superconducting materials; mixed-valent compounds and heavy fermions; interfaces and buried layers; More generally, rare earth and transition metal compounds, especially oxides. >> Resonant IXS, Resonant HAXPES, Auger spectroscopy. | | Chemistry | | Valence properties and local structure in metallic complexes; in-situ chemical analysis, catalysis, electrochemistry. >> High resolution x-ray absorption spectroscopy, X-ray fluorescence and resonant emission. | | Geophysics, biophysics | | Magnetism under high pressure; K-edge of light elements in extreme conditions. >> Fluorescence, X-ray Raman spectroscopy. |
|
Contacts
|
|
|
Employment
|
|